Customization: | Available |
---|---|
Classification: | Imaging Diagnostic Equipment |
Type: | Ultrasound Scanner |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
.12.1 inch display
.backlight keys
.Built-in battery(capacity 3000mAh)
.Thin and light notebook design
.16 kinds pseudo color
.Color interface
.Built-in date report
laptop Full Digital Ultrasound Diagnostic Scanner Equipment
Main Functions:
---Scanning mode: electron scanning
---support Chinese and English display
---Zoom: Local x 2.0 magnification rate
---Dynamic range: 0~120dB adjustable.
---Focus position: Send: 1, 2, 3 and 4-segments dynamic electron focusing; when 1,2-segments focus,
the focus can be moved.
---Receive: receive real-time dynamic focus by focus
---Measuring function: Distance, circumference/area (method of ellipse, method of loci), volume, heart rate, gestational weeks (BPD, GS, CRL, FL, HC,AC), expected date of confinement and fetus weight, etc.
---Annotation function: hospital name, patient's name, age and gender;
---64 body marks (with probe's position); Full-screen character annotation; Real-time clock display;
---Puncture guide: 3.5MHz convex array probe can display puncture guide line in B mode.
---Gain control: 8 segments TGC and overall gain can be adjusted respectively.
---Image polarity: left and right flip, black and white flip, up and down flip.
---Permanent storage: >=128 images.
---Image playback: series playback or check one by one.
---Image processing: γ correction, frame correlation, point correlation, line correlation, digital filtering, digital edge enhancement and pseudo color processing, etc.
---USB interface: store image only for U disk.
Main Technical Indexes:
Standard probe: T35R60BN convex array probe: 80 elements, R60, frequency of 3.5 MHz
Optional probe: T65R13BN type cavity probe: 80 elements, R13, frequency of 6.5 MHz
Basic performance as below:
NO. |
Name of technical indexes |
Type T35R60BN convex array probe |
Type T65R13BN cavity probe |
1 |
Nominal frequency of probe MHz |
f=3.5 MHz |
f=6.5 MHz |
2 |
Type of probe |
R=60,Convex array |
R=13,Micro-convex array |
3 |
Scan depth mm |
≥170 mm |
≥60 mm |
4 |
Lateral resolution mm |
≤3mm(depth≤80 mm) |
≤2mm(depth≤30 mm) |
≤4mm(80 mm |
|||
5 |
Axial resolution mm |
≤2mm(depth≤80 mm) |
≤1mm(depth≤40 mm) |
≤3mm(80 mm |
|||
6 |
Blind area |
≤5 mm |
≤4 mm |
7 |
Transverse geometry precision |
≤10 % |
≤10 % |
8 |
Longitudinal geometric location accuracy |
≤8 % |
≤5 % |
---Monitor: 12.1inch LED display.
---Grey scale: 256 levels.
---The perimeter area measurement deviation: 20% or less.
---Running hours: ≥8h.
---The battery pack continuous working time: ≥1h.
---Input power: ≤110VA.
---Host weight: about 3.5 kg.
---Host appearance size: 310 ×338 × 76 (length × width × height) (mm3).
Configuration probe(Optional)
The configuration probe of the instrument showed as below:
Type |
Scan model |
Frequency range |
element |
radius of curvature |
Scan width |
Scan angle |
T35R60BN |
Convex |
2.5MHz~5.0MHz |
80 |
R60mm |
_ |
60° |
T75L40BN |
Linear |
5.0MHz~7.0MHz |
80 |
_ |
L40mm |
_ |
T65R13BN |
Cavity |
5.5MHz~7.5MHz |
80 |
R10mm |
_ |
150° |
T50R20BN |
Micro-C |
5.0MHz |
80 |
_ |
_ |
90° |
Configuration
Standard configuration: 1 host; 1 T35R60BN convex array probe.
Optional configuration: T65R13BN type cavity probe.